Vivid S5 N / Vivid S6 N Battery Safety
NOTE:
The Vivid S6 N ultrasound scanner is supplied with a lithium ion battery in the battery bay, as standard.
For the Vivid S5 N, it is supplied as an option.
The lithium ion battery provides power to safely shutdown the system or place it in Standby mode, when
an AC power source is not available. Lithium ion batteries last longer than conventional batteries and
do not require replacement as often. You can expect 4 hours of battery life with a single, fully-charged
battery supporting Standby mode.
The lithium ion technology used in the system's battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such a watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:
Regulations vary for different countries. Dispose of a used battery in accordance with local regulations.
Use only batteries approved by GE HEALTHCARE as suitable for use with the Vivid S5 N / Vivid
CAUTION
S6 N ultrasound scanner
THE VIVID S5 N / VIVID S6 N BATTERY HAS A SAFETY DEVICE.
WARNING
WARNING
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
Charge batteries only when the ambient temperature is between 0
and discharge the batteries between -10
Do not short-circuit the battery by directly connecting the negative terminals with metal objects.
Do not heat the battery or discard it in a fire.
Do not expose the battery to temperatures over 60
other heat sources.
Do not charge the battery near a heat source, such as, a fire or heater.
Do not leave the battery in direct sunlight.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery.
Do not apply solder to a battery.
Do not connect the battery to an electrical power outlet.
In the event that the Vivid S5 N or Vivid S6 N ultrasound scanner will not be used for a
WARNING
WARNING
long period, it is recommended to remove the battery from the system while not in use.
To prevent the battery bursting, igniting, or fumes from the battery causing equipment damage,
CAUTION
always observe the following precautions:
Do not immerse the battery in water or allow it to get wet.
Do not place the battery into a microwave oven or pressurized container.
If the battery leaks or emits an odor, remove it from all possible flammable sources.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
If you have any questions about the battery, consult your local GE representative.
1-14
2
EVISION
Section 1-3 - Safety Considerations
VS5 N
o
o
o
and 55
C (14
and 131
o
o
C (140
F). Keep the battery away from fire and
VS6 N
S
AND
ERVICE
o
o
o
and 65
C (32
and 149
o
F)
M
ANUAL
o
F).