GE Vivid S5 N Service Manual page 619

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-4-15
Speaker Replacement Procedure
NOTICE
8-4-15-1
Tools
Use the appropriate flat screwdriver and T10 Torx screwdriver (T8 may also be used), as indicated in
the Speaker replacement procedure.
8-4-15-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-4-15-3
Speaker Removal Procedure
1.) Remove the Keyboard Assembly, as described in the
page
2.) Carefully place the Keyboard Assembly upside-down on a flat, clean, stable surface.
3.) Remove the Keyboard Bottom Cover, as described in the
Procedure" on page
4.) Disconnect the two cables from the Extended Keyboard, as indicated in
2
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
"Battery Removal Procedure" on page
8-39.
8-59.
C
C
Note:
= cable to be disconnected
Figure 8-106 Cables Connected to the Extended Keyboard
Chapter 8 - Replacement Procedures
VS5 N
VS6 N
AND
8-116.
4-2-3 "Power Shut Down" on
"Keyboard Assembly Removal Procedure" on
"Keyboard Bottom Cover Removal
Figure
C
S
M
ERVICE
ANUAL
8-106.
Extended Keyboard
8-93