Download Print this page

GE Vivid S5 N Service Manual page 626

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-5-1-3-3
Removing Cables from the Upper Arm Assembly
Note:
After removing cables from the Keyboard Interface Assembly, it is recommended to lean
the keyboard interface platform backwards and to place protective material between the
lower arm and the handle, as indicated in
tie the lower arm and handle together in this position, using string or a long tie wrap, as
appropriate.
Place Protective
Material Here
1.) From below the Keyboard Interface Assembly, remove the two upper cable securing brackets, as
indicated in
(Use a Phillips screwdriver to carefully loosen the screw and remove it, together with the bracket
Bracket
8-100
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
Figure 8-117 Positioning the Keyboard Interface Platform
Figure
8-118.
Figure 8-118 Upper Cable Securing Brackets
VS5 N
Figure
8-117. It is also advisable to temporarily
VS6 N
S
M
AND
ERVICE
ANUAL
Bracket
).

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n