Gas Spring Replacement Procedure
8-7-4-1
Tools
Use the appropriate flat and Phillips type screwdrivers and the appropriate Allen keys.
8-7-4-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-7-4-3
Gas Spring Removal Procedure
1) Remove the monitor, as described in the
2) Remove Keyboard Assembly as described in the
page
3.) Remove Lower Arm as described in the
4.) If the rear tray is attached to the system, remove it.
5.) Using a flat screw driver, carefully remove the plastic covering caps from the support arm, as shown
in
8-198
2
EVISION
8-39.
Figure
8-249.
Figure 8-249 Removing Protective Plastic Caps
Section 8-7 - Mechanical Components Replacement Procedures
"Monitor Removal Procedure" on page
"Keyboard Assembly Removal Procedure" on
"Lower Arm Removal Procedure" on page