Control Console Components Replacement Procedures

8-3-1
Monitor Replacement Procedure
NOTE:
This section describes the procedure for replacing either a 15" or 17" monitor, as applicable
(see details below):
NOTE:
For monitor operation and image quality calibration instructions for both the 15" and the
17" monitor, refer to the information provided in
8-3-1-1
Tools
Use the appropriate Phillips screwdriver, as indicated in the monitor replacement procedures.
8-3-1-2
Preparation
Shut down the Vivid S5 N ultrasound unit, as described in
8-3-1-3
Monitor Removal Procedure
1) Working from the front of the system, lower the console to the maximum down position.
2) Make sure that the console is aligned in the central position (not pulled to one side or the other).
3) Working from the left or right of the support column, pull the locking pin slightly out to release the
monitor from the locked upright position, as shown in
8-26
Section 8-3 - Control Console Components Replacement Procedures
2
EVISION
Vivid S5 N systems: 15" monitor
Vivid S6 N systems: 17" monitor
Figure 8-27 Releasing the Monitor from the Locked Upright Position
VS5 N
Chapter 6 - Service
Adjustments.
4-2-3 "Power Shut Down" on page
Figure 8-27
below.
VS6 N
S
M
AND
ERVICE
4-7.
Locking Pin
ANUAL