Cable Replacement Procedures

8-5-1
Keyboard and Monitor Cable Replacement Procedure
The Vivid™ S5 N and Vivid™ S6 N system's communication and power cables extend from the LCD
monitor downwards along the Monitor Arm Assembly, to the Keyboard Interface Assembly, and finally
to the Distribution Interface Board (DIB). The replacement procedures are carried out in stages, as
described in the sub-sections below.
The following stages show the order in which cables are removed/replaced:
LCD monitor
Keyboard assembly
Upper/Lower Arm Assembly
Keyboard Interface Assembly
DIB
8-5-1-1
Tools
Use the appropriate flat and Phillips screwdrivers, Allen Key, appropriate flat spanner, and diagonal side
cutter as indicated in the Keyboard or Monitor Cable replacement procedure.
8-5-1-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-5-1-3
Keyboard and Monitor Cable Removal Procedure
1.) Remove the Monitor, as described in
2.) Remove the three Allen screws from the LDC metal mounting plate, as show in
8-5-1-3-1
Removing Cables from the Monitor Arm Assembly
Figure 8-108 Screws Securing Mounting Plate on the Support Column
2
EVISION
Chapter 8 - Replacement Procedures
"Monitor Removal Procedure" on page
VS5 N
VS6 N
S
AND
ERVICE
4-2-3 "Power Shut Down" on
8-26.
Figure
8-108.
Mounting Plate
M
ANUAL
8-95