GE
P
N
FN091065, R
ART
UMBER
11.)Using a screwdriver, tap out the supporting pins to release the Joint Assembly -
12.)Release the communication cables from the arm.
13.)Remove the cables by passing them through the Joint Assembly aperture.
14.)Remove the Joint Assembly.
8-4-3-4
Joint Assembly Installation Procedure
1.) Return the Joint Assembly to its location on the Upper Arm and insert the supporting pins.
2.) Secure the support pins by tightening the two locking screws -
3.) Return the ring clips to their positions on the supporting pins.
4.) Return the communication cables by passing them through the Joint Assembly aperture.
5.) Return the gas spring cable to its position (see
must face upwards -
6.) Reconnect the Ground lead.
7.) Return the Keyboard Interface Assembly to its position as described in the
Assembly Installation Procedure" on page
8.) Install the Lower Arm as described in the
9.) Install the Probe Shelf as described in the
10.)Install the Keyboard Assembly as described in the
page
11.)Return the Monitor to its position as described in the
2
EVISION
Figure 8-83 Removing the Joint Assembly
Figure
8-84.
Figure 8-84 Gas Spring Cable Bracket showing Cable Slot
8-42.
Chapter 8 - Replacement Procedures
Tap out
Supporting
Pin
Figure
8-80) with its bracket - the slot in the bracket
8-47.
"Lower Arm Installation Procedure" on page
"Probe Shelf Installation Procedure" on page 8-52
"Keyboard Assembly Installation Procedure" on
"Monitor Installation Procedure" on page
VS5 N
VS6 N
S
AND
ERVICE
Figure
Figure
8-81.
Slot in Bracket Facing Upwards
"Keyboard Interface
8-196.
M
ANUAL
8-83.
8-29.
8-67