PIB (Probe Interface Board) Board Replacement Procedure
6.) Disconnect the flex cable from the PIB Board.
7.) Loosen and remove the five screws holding the PIB Board in position (see
NOTE:
There are 3 long and 2 short screws. (The other short screw seen in
removed when disconnecting the PIB-to-PSB flex cable.
8.) Remove the PIB Board from the cage assembly.
8-154
2
EVISION
Figure 8-189 Removing the PIB Board
Figure 8-190 PIB Board Removed