P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-5-1
Accessing the Monitoring Diagnostic Test Options
Figure 7-60 View Test Log - Full System Monitoring Test - Fixed Voltages
9.) Check that the sub-test passed and that result values are within the pre-defined Max and Min limits,
as shown in
92.
10.)Select the second report (Temperatures); the report is highlighted in the Reports list, and the
corresponding results are shown in the Report window below, as shown in
Log - Full System Monitoring Test - Temperatures" on page
7-92
2
EVISION
Figure 7-60 "View Test Log - Full System Monitoring Test - Fixed Voltages" on page 7-
Section 7-5 - Monitoring Tests
VS5 N
VS6 N
S
AND
(cont'd)
Figure 7-61 "View Test
7-93.
M
ERVICE
ANUAL