P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-6-1
Accessing the Back End Diagnostic Test Options
7-6-1-1-7 PCI Test
NOTE:
For a description of this test, refer to the information in
1.) In the Diagnostic Test window, trackball to PCI Test in the list of PC Doctor tests available for selection
from the Diagnostic Test Tree (see
name is displayed under Execute; the relevant system components to be tested are highlighted light
blue in the Schematic View (BEP, Peripherals, Keyboard), as shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
3.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the Start button and press Set.
The PCI Test commences. While the test proceeds, the name of the test currently in progress is
displayed in the Execute field and the current Loop # and specific test are shown below. The
messages Init Done and Please Wait are displayed beside the progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
7-128
2
EVISION
Green: All recommended tests for these components have been completed and no problems
were identified.
Red:
Problems were identified in these components during the performance of the test.
Section 7-6 - Performing Back End Diagnostics on the System
"PCI Test" on page
Figure 7-89 "PCI Test" on page
Figure 7-89 PCI Test
VS5 N
VS6 N
S
AND
ERVICE
(cont'd)
7-19.
7-128) and press Set. The Test