P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-7-3
Lower Arm Replacement Procedure
8-7-3-1
Tools
Use the appropriate flat and Phillips-type screwdrivers as indicated in the Lower Arm replacement
procedure.
8-7-3-2
Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-7-3-3
Lower Arm Removal Procedure
1) Working from the front of the system, raise the console to the maximum up position.
2) Make sure that the console is aligned in the central position (not pulled to one side or the other).
3.) Loosen and remove the screw that is located at the base of the operator panel (below the
keyboard), as shown in
2
EVISION
Figure 8-242
Figure 8-242 Removing Screw on Left Side of Lower Arm
Chapter 8 - Replacement Procedures
below.
VS5 N
VS6 N
S
AND
ERVICE
4-2-3 "Power Shut Down" on
M
ANUAL
8-193