P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Table 7-3
Option
Reports
Reports Scroll
Bar and Arrows
Erase Button
Exit Button
Report Window Lower portion of the View Log dialog box where the
Report window
Scroll Arrows
Close Button
7-6
2
EVISION
Vivid™ S5 N and Vivid™ S6 N Diagnostics Test Window - View Log (cont'd) 2 of 2
Description
Displays the specific report (or list of reports) for
the currently-selected test sequence.
(shows all OK and failed tests)
Certain tests will have more than one report (see
the example in Figure 7-2 - View Log Dialog Box).
Enables scrolling through the list of Reports
Clears all test sequence log files for that specific
(currently selected) test listed in the Tests
window (passed and failed tests).
Closes the View Log dialog box
the currently-selected report is displayed in a
separate window (see the example in
"View Log Dialog Box" on page
The displayed Average, Deviation, Max and Min
values are dynamically updated.
Enables scrolling through all channels.
Displays the currently-selected channel number.
For example, Count: 64
Closes the Report window.
Section 7-2 - Diagnostics
VS5 N
Double-clicking on a currently-selected
report displays that report in the Report
window (lower portion of the View Log
dialog box) - refer to
- Viewing Report (Test Signal Amplitude
2nd Harmonic)" on page
A test sequence may have up to four
reports. Since only one specific test may
have failed, the scroll arrows enable
jumping directly from one Failed test
report to the next (i.e., failures only).
Alternatively, you can scroll through all the
reports (one-by-one).
Jump to Failed Test Reports
It is not recommended to clear the test
sequence log files unless a test failed due
to user intervention or as a result of a false
alarm. (It is necessary to erase each log
file separately).
Returns to the Diagnostic Test window.
Visible only after test execution.
Figure 7-2
7-7).
VS6 N
S
M
AND
ERVICE
ANUAL
Comments
Figure 7-3 "ADC Test
7-8.
Scroll between
reports here