PIB (Probe Interface Board) Board Replacement Procedure
NOTICE
8-6-9-1
Tools
Use the appropriate flat and Phillips-type screwdrivers as indicated in the PIB replacement procedure.
8-6-9-2
Time Required
10 minutes
8-6-9-3
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-6-9-4
PIB Board Removal Procedure
1) Remove the system left and right side covers, as described in the
Procedure" on page 8-6
2) Remove the system front cover, as described in the
3) Remove the cabinet cage assembly from the system, open the cage and carefully place it on a
stable surface, as described in the
page
4.) Unfold the cabinet cage assembly, as described in
Assembly" on page 8-109
5) Remove the screw that secures the PIB-to-PSB flex cable in position
2
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
"Battery Removal Procedure" on page
and
"Right Side Cover Removal Procedure" on page
8-109.
(see
Figure 8-188 Unscrewing the PIB-to-PSB Flex Cable
Chapter 8 - Replacement Procedures
"Front Cover Removal Procedure" on page
"Cabinet Cage Assembly Opening and Removal Procedure" on
"Opening and Removing the Cabinet Cage
step 11
on page 8-112).
VS5 N
VS6 N
AND
8-116.
4-2-3 "Power Shut Down" on
"Left Side Cover Removal
8-5.
(Figure
8-188).
S
M
ERVICE
ANUAL
8-17.
8-153