P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-4-6
Probe Select Board Diagnostic Tests
The ACW Matrix Test commences. While the test proceeds, the name of the test currently in
progress (ACW IQ Symmetry Test) is displayed in the Execute field and the current Loop # and
specific test are shown below. The messages Init Done and Please Wait are displayed beside the
progress bar in the Status area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the ACW Matrix Test sequence is complete:
-
-
2
EVISION
Green: All recommended tests for these components have been completed and no problems
were identified.
Red:
Problems were identified in these components during the performance of the test.
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
Matrix Test - Signal Frequency" on page
Figure 7-39 View Test Log - ACW Matrix Test - Signal Frequency
Chapter 7 - Diagnostics/Troubleshooting
VS5 N
(cont'd)
Figure 7-39 "View Test Log - ACW
7-67.
VS6 N
S
M
AND
ERVICE
ANUAL
Expected Average
Selected channel
Average Result
Scroll between
channels here
Selected channel # (32)
7-67