N
FN091065, R
ART
UMBER
8.) Remove the two screws that secure the console holder, then remove the holder
9.) Carefully remove the adhesive tape from the top of the system and spread the anti-static nylon on
the floor.
10.)Remove any additional packing material from in front of and surrounding the system.
11.)Remove the two silica gel bags
12.)Press down on the locking brake to unlock the wheel. Repeat for the three other wheels.
13.)Using two people, carefully roll the scanner forwards down the unloading ramp to remove it
completely from the shipping carton.
CAUTION
14.)Make sure the system is standing firmly on the floor before applying the locking brake.
15.)Remove the foam sections from the top of the system and from between the main shelf and
probe shelf.
16.)Using the up/down handle, raise the main shelf up to appropriate height.
17.)Pull the monitor locking pin towards you and lift the monitor into the upright position.
18.)Before commencing the installation procedure, follow the instructions for
Carton Contents" on page
3-12
2
EVISION
Locking brake
Silica gel bags
Figure 3-9 Locking Wheel Brake
TWO PEOPLE ARE REQUIRED WHEN MOVING THE Vivid S5 N or Vivid S6 N
SCANNER.
This precaution is intended to prevent injury that may be caused by the weight of the
system if one person attempts to move the scanner considerable distances or on an
incline.
3-13.
(Figure
3-9).
-
VS5 N
VS6 N
S
AND
ERVICE
(Figure
Anti-static nylon on floor
'Verifying the Shipping
M
ANUAL
3-7).