P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-4-6
Probe Select Board Diagnostic Tests
7-4-6-2-3 ACW Delay Range Test
NOTE:
For a description of this test, refer to the information in
In order to access an individual board test, it is necessary to insert the service dongle from system boot-
NOTICE
up and enter the Service Password.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be touching
the probe connectors!
It is necessary to disable the Loop option where it should not be used.
NOTICE
After performing this test, you must re-boot the system and perform the DC Offset Calibration procedure
as described in
1.) In the Diagnostic Test window, trackball to ACW Delay Range Test in the list of ACW tests
available for selection from the Diagnostic Test Tree (see
page
relevant system components to be tested are highlighted light blue in the Schematic View (PIB &
PSB), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Adjacent to On Fail, select either Stop or Continue, as required. Stopping on failure enables you
to review a failed test, the moment it fails. Alternatively, you may prefer testing to continue on failure
and later you may review the report.
2
EVISION
"DC Offset Calibration" on page
7-71) and press Set. The ACW Delay Range Test name is displayed under Execute; the
Figure 7-42 ACW Delay Range Test
Chapter 7 - Diagnostics/Troubleshooting
VS5 N
(cont'd)
"ACW Delay Range Test" on page
7-27.
Figure 7-42 "ACW Delay Range Test" on
VS6 N
S
M
AND
ERVICE
ANUAL
7-14.
7-71