GE Vivid S5 N Service Manual page 645

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-6-3
Battery Bracket Replacement Procedure
NOTE:
This procedure is applicable to Vivid S6 N systems only.
8-6-3-1
Tools
Use the appropriate Phillips-type screwdrivers and a side cutter as indicated in the Battery Bracket
replacement procedure.
8-6-3-2
Time Required
1 minute
8-6-3-3
Preparations
Shut down the Vivid S6 N ultrasound unit, as described in
8-6-3-4
Battery Bracket Removal Procedure
1) Remove the Vivid S6 N left and right side covers, as described in the
Removal Procedure" on page 8-6
2) Remove the Vivid S6 N front cover, as described in the
on page
3) Open and remove the cabinet cage assembly and place it on a stable surface, as described in
1 "Opening and Removing the Cabinet Cage Assembly" on page
4.) Remove the battery, as described in the
5.) Cut the cable tie wraps.
6.) Loosen (do not remove) the four screws that secure the battery bracket to the cage assembly, as
indicated in
7.) Carefully lift the end of the battery bracket that is facing you upwards (while sliding it forwards
slightly to free it from the secured position) to partially remove it from the cage
2
EVISION
8-17.
Figure
8-147.
Figure 8-147 Loosening the Battery Bracket Securing Screws
Chapter 8 - Replacement Procedures
4-2-3 "Power Shut Down" on page
and
8-2-2-3 "Right Side Cover Removal Procedure" on page
8-2-6-3 "Front Cover Removal Procedure"
8-6-2-4 "Battery Removal Procedure" on page
VS5 N
VS6 N
S
AND
ERVICE
8-2-3-3 "Left Side Cover
8-109.
8-116.
(Figure
8-148).
M
ANUAL
4-7.
8-5.
8-6-