Table 8-2 Step-by-Step Software Upgrade/Installation/Reinstallation Procedure (cont'd) 9 of 9
Step
18.
Verify the Application Software version corresponds
with the System Software version you have installed
(check against the label on the USB Flash Drive) and
that it was correctly installed.
19.
Restore Patient Archive and System Configuration
as per User Manual.
Note: When performing "Restore", select "System
Configuration" and then de-select "Imaging Preset" and
"GSP Resources".
20.
Install peripherals.
21.
Restore the settings as recorded under Preparation
section 8-8-2
on page 8-209.
Verify proper Connectivity functionality - refer to
"Connectivity Setup" on page
22.
Proceed to perform the following functionality tests:
-
"Software Configuration Checks" on page 4-45
-
"Calibration" on page 7-12
-
"Full System Test" on page 7-11
2
EVISION
Instruction
3-96.
Chapter 8 - Replacement Procedures
VS5 N
VS6 N
AND
Expected Result
Deselect Imaging Presets
Deselect GSP Resources