Download Print this page

GE Vivid S5 N Service Manual page 357

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
5-6-3
AC Distribution Panel
The distribution of isolated local AC power is allocated in the peripherals section, and divided as follows:
A single Isolated Internal power connection for optional integrated BW Printer
Two Isolated External power outlets for optional connection of external peripherals
2 isolated AC outlets
Ground Equalization
2
EVISION
Figure 5-30 Location of AC Distribution Panel
Table 5-17 AC Distribution Panel - Connectors
Connector #
P1
P2
J6
GB#1
Connector - J6
Figure 5-31 Location of AC Distribution Box J6 Connector
Chapter 5 - Components and Function (Theory)
Circuit Breaker
AC Input
Description
AC input to system
Isolated AC input
Isolated DC output
System Ground Equalization
VS5 N
VS6 N
S
AND
ERVICE
M
ANUAL
5-39

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n