P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Note:
When reconnecting the ground leads, there are two spring washers that should be installed in
the order shown in
surface facing towards the Cart to prevent damage to the cable holder and the cable.
Figure 8-237 Ground Lead Showing Spring Washers and Cable Holder
10.)Return the bottom rear cover to the system, as described in
Procedure" on page
11.)Turn ON power to the system.
12.)Proceed to perform the following functionality tests:
-
-
-
-
8-188
2
EVISION
Figure
8-237. In addition, place the ground lead connector with the flat
8-23.
"Grounding Continuity" on page 10-23
"Chassis Current Leakage Test" on page 10-24
"Calibration" on page 7-12
"Full System Test" on page 7-11
Section 8-6 - Lower Section Components Replacement Procedures
VS5 N
VS6 N
S
AND
Ground Lead
"Bottom Rear Cover Installation
M