Download Print this page

GE Vivid S5 N Service Manual page 579

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-3-7
Swivel and Up-Down Handle Replacement Procedure
8-3-7-1
Tools
Flat and Phillips screwdrivers as needed.
8-3-7-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-3-7-3
Swivel and Up-Down Handle Replacement Procedure
1) Remove the Operator Panel as described in the
page
2) Remove the Probe Shelf as described in the
3.) Unscrew the protective metal plate (shown in
Swivel Cable
Up-down Cable
2
EVISION
8-39.
Figure 8-66 Protective Metal Plate Rotated 90 Degrees
Chapter 8 - Replacement Procedures
VS5 N
4-2-3 "Power Shut Down" on
"Keyboard Assembly Removal Procedure" on
"Probe Shelf Removal Procedure" on page
Figure
8-66) and rotate it upwards 90 degrees.
VS6 N
S
M
AND
ERVICE
ANUAL
8-52.
8-53

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n