Download Print this page

GE Vivid S5 N Service Manual page 95

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Table 3-5
2
EVISION
Damage Inspection Checklist - Vivid S5 N and Vivid S6 N Systems (cont'd)
Step
Item
16
Power Cord
17
18
Front Castors
19
Rear Castors
Chapter 3 - System Setup
Recommended Procedure
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or
any deterioration. Verify continuity.
Tighten the clamps that secure the power cord to the unit and the outlet plug to the
cord. Replace the power cord and/or clamp, as required.
Secure the cable to the back of the system using the plastic clamp provided.
Check that the front castors can roll and swivel, and can be placed in the locked
position by pressing the foot brake (lower lever) down on each.
Ensure that the wheels are locked and there is no movement in any direction.
For Vivid S6 N systems:
Check that the rear castors can roll and swivel, and can be placed in the locked
position by pressing the foot brake (lower lever) down on each.
Ensure that the wheels are locked, unable to swivel left or right, and that there is
no movement forwards or backwards.
For Vivid S5 N systems:
Check that the rear castors can roll and swivel freely.
VS5 N
VS6 N
S
AND
ERVICE
Clamp securing power cord
AC cable secured with plastic clamp
M
ANUAL
3-19