ART
UMBER
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
2
EVISION
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
"DC Offset Calibration" on page
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
Chapter 7 - Diagnostics/Troubleshooting
7-27, and then repeat the test.
(Theory).
VS5 N
VS6 N
S
AND
ERVICE
Chapter 5 -
M
ANUAL
7-89