Download Print this page

GE Vivid S5 N Service Manual page 504

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
7-6-1
Accessing the Back End Diagnostic Test Options
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the CPU Full Test sequence is complete:
-
-
NOTE:
For the CPU Full Test, many tests are performed in the test sequence; these are listed in the
Tests window as shown in
Registers) the corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
-
-
-
-
-
-
-
-
-
-
-
-
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
7-120
2
EVISION
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in
Test Log - Full Back End Test w/o Hard Disk Scan" on page
Figure 7-82 View Test Log - CPU Full Test
Figure
CPU Registers
CPU Arithmetics
CPU String Operations
CPU Logical Operations
CPU Math Registers
CPU Math Commands
CPU Transcendental
CPU MMX Arithmetic
CPU MMX Logical
CPU MMX Shift
CPU MMX Data Transfer
CPU MMX Misc.
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function
Section 7-6 - Performing Back End Diagnostics on the System
7-80. For each currently-selected test (for example,
(Theory).
VS5 N
VS6 N
S
AND
ERVICE
(cont'd)
Figure 7-78 "View
7-116.
Chapter 5 -
M
ANUAL

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n