Download Print this page

GE Vivid S5 N Service Manual page 299

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
Table 4-12
Step
5
6
1
2
3
4
5
6
2
EVISION
Front End Tests (cont'd)
Task
Trackball to the Start button and press Set
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
ACW Matrix Test
Trackball to ACW Matrix Test and press Set.
Disconnect any probes currently connected to the
system and make sure that nothing is touching the probe
connectors.
Under the Special Instructions area, activate the Loops
checkbox and select the required number of Loops from
the combo box.
Adjacent to On Fail, select either Stop or Continue, as
required.
Trackball to the Start button and press Set
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
Chapter 4 - Functional Checks
VS5 N
VS6 N
AND
Expected Result(s)
The test commences.
When the ACW IQ Symmetry Test sequence is
complete, the View Test Log window opens
automatically, listing the test performed and showing
whether it passed or failed - see the example in
7-37
on page 7-65.
The test commences.
When the ACW Matrix Test sequence is complete, the
View Test Log window opens automatically, listing the
test performed and showing whether it passed or failed
- see the examples in
Figure 7-39
Figure 7-40
on page 7-69.
S
M
ERVICE
ANUAL
Figure
on page 7-67 and
4-29

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Vivid s6 n