Download Print this page

GE Vivid S5 N Service Manual page 595

Hide thumbs Also See for Vivid S5 N:

Advertisement

P R E L I M I N A R Y
GE
P
N
FN091065, R
ART
UMBER
8-4-4
Alphanumeric Keyboard Replacement Procedure
NOTICE
8-4-4-1
Tools
Use the appropriate flat screwdriver and T10 Torx screwdriver (T8 may also be used), as indicated in
the Alphanumeric Keyboard replacement procedure.
8-4-4-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-4-4-3
Alphanumeric Keyboard Removal Procedure
1.) Remove the Keyboard Assembly, as described in the
page
2.) Carefully place the Keyboard Assembly upside-down on a flat, clean, stable surface.
3.) Remove the Keyboard Bottom Cover, as described in the
Procedure" on page
4.) Loosen the securing screw from each of the two GND cables and release the cables (see
Figure
C
Note:
= cable to be disconnected
2
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
"Battery Removal Procedure" on page
8-39.
8-59.
8-85).
C
C
Alphanumeric Keyboard
Figure 8-85 Removing the Alphanumeric Keyboard
Chapter 8 - Replacement Procedures
VS5 N
VS6 N
AND
8-116.
4-2-3 "Power Shut Down" on
"Keyboard Assembly Removal Procedure" on
"Keyboard Bottom Cover Removal
C
C
C
C
GND Cable
S
M
ERVICE
ANUAL
GND Cable
8-69

Hide quick links:

Advertisement