Keyboard Assembly Replacement Procedure
8-3-4-1
Tools
Use the appropriate flat and Phillips-type screwdrivers and the appropriate Hex wrench keys, as
indicated in the keyboard assembly replacement procedure.
8-3-4-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-3-4-3
Keyboard Assembly Removal Procedure
1) Working from the front of the system, raise the console to the maximum up position.
2) Make sure that the console is aligned in the central position (not pulled to one side or the other).
3) Working from beneath the keyboard assembly, loosen and remove the two Allen screws on the right
side, as shown in
Figure 8-47 Removing the Allen Screws under Keyboard Assembly - Right Side
NOTE:
Adjacent to each of the four Allen screws (two on the left; two on the right) there is an arrow
marking the position of the screws for easy identification:
2
EVISION
Figure 8-47
below (arrows marked in the figure indicate the screw locations).
Chapter 8 - Replacement Procedures
VS5 N
VS6 N
S
AND
ERVICE
4-2-3 "Power Shut Down" on
M
ANUAL