TR32 and RFI Boards Replacement Procedure
7) Loosen and remove the 10 Phillips screws that fasten the RFI board in position -
NOTE:
There are 2 long, 2 medium length, 2 short, and 4 very short screws.
8) Disconnect the power cable from the RFI Board - see
9) Disconnect the BEP-to-RFI flex cable from the cable connector on RFI Board, as shown in
8-193
BEP-to-RFI
Flex Cable
2
EVISION
Figure 8-192 Removing Securing Screws from RFI Board
below.
Figure 8-193 Disconnecting the Cables from the RFI Board