ECG Module Assembly Replacement Procedure
NOTICE
8-6-6-1
Tools
Use the appropriate flat screwdriver, as indicated in the ECG Module Assembly replacement procedure.
8-6-6-2
Time Required
10 minutes
8-6-6-3
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
page
4-7.
8-6-6-4
ECG Module Assembly Removal Procedure
1) Remove the system left and right side covers, as described in the
Procedure" on page 8-6
2) Remove the system front cover, as described in
8-17.
3) Open and remove the cabinet cage assembly and place it on a stable surface, as described in
1 "Opening and Removing the Cabinet Cage Assembly" on page
NOTE:
The location of the ECG Module Assembly is shown in
.
ECG module
8-140
2
EVISION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
"Battery Removal Procedure" on page
and
Figure 8-170 ECG Module Assembly
Section 8-6 - Lower Section Components Replacement Procedures
8-2-2-3 "Right Side Cover Removal Procedure" on page
8-2-6-3 "Front Cover Removal Procedure" on page
Figure 8-170
VS5 N
VS6 N
AND
8-116.
4-2-3 "Power Shut Down" on
8-2-3-3 "Left Side Cover Removal
8-109.
USB cable
ECG cable
S
M
ERVICE
ANUAL
8-5.
8-6-