Display Assembly Replacement Procedures

5-20
Maintenance: Display Assembly Replacement Procedures
1.
Remove the battery as described in
2.
Remove the two fasteners that secure the display panel cover at the back of the
unit and then remove the display cover.
3.
Remove the two fasteners holding the two ground wires on either side of the
display.
4.
Pivot the display up to access the cable connections on the main PCB.
NOTE
If a KISS pump is installed, it must be removed to access the display panel
connectors from the main board. See
Procedures".
5.
Disconnect the blue ribbon cable from the main PCB board.
6.
Disconnect the display light cable from the main PCB board.
7.
Push back on each of the tabs holding the display assembly to the pivot as
shown below. Gently pull the display assembly free.
8.
Reverse disassembly procedures with a new display assembly.
MAC™ 3500 Resting ECG Analysis System
2021337-036
"Battery Replacement"
on page
"KISS Pump Replacement
95A
5-6.