Accessing the Back End Diagnostic Test Options
5.) At any stage, trackball to the Pause button and press Set, if required.
6.) To resume testing (from the point where you paused), trackball to the Start button and press Set.
When the Hard Drive Full Test w/o Scan sequence is complete:
-
-
NOTE:
For the Hard Drive Full Test w/o Scan, several tests are performed in the test sequence; these
are listed in the Tests window as shown in
o Scan" on page
corresponding report is listed in Reports on the right.
7.) Use the scroll bars and/or arrows to scroll through the list of Test results and review the reports, as
required:
-
-
-
8) When done, trackball to the Exit button and press Set to close the View Report window.
9) After reviewing the test results, proceed as follows:
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
2
EVISION
Yellow: The tests executed so far on these components have passed, but not all
recommended tests have been performed.
the Diagnostic Test window displays the Finished message
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests and sub-tests passed or failed, as shown in the example in
Test Log - Hard Drive Full Test w/o Scan" on page
Figure 7-84 View Test Log - Hard Drive Full Test w/o Scan
7-123. For each currently-selected test (for example, Linear Seek) the
Linear Seek
Random Seek
Funnel Scan
If no problems were identified, it is recommended to re-boot the system (this is obligatory when
prompted).
If no problems were identified but the problem persists, contact your local GE HEALTHCARE
field engineer for assistance.
If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in
Components and Function
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
Chapter 7 - Diagnostics/Troubleshooting
Figure 7-84 "View Test Log - Hard Drive Full Test w/
(Theory).
VS5 N
VS6 N
AND
(cont'd)
Figure 7-84 "View
7-123.
S
M
ERVICE
ANUAL
Chapter 5 -
7-123