P R E L I M I N A R Y
D
FR091521, R
IRECTION
Section 8-3
Control Console Components - Replacement Procedures
8-3-1
21.5" Monitor Replacement Procedure
8-3-1-1
Tools
Phillips screwdriver
FRU Part #
Refer to
8-3-1-2
Time Required
15 min
8-3-1-3
Preparation
1.) Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
4 -
7.
1) Lower the console to the maximum down position.
Make sure it is aligned in the central position (not pulled to one side or the other).
2.) Fold the monitor forward into the face-down position, then lock the articulated arm.
8-3-1-4
Remove the 21.5 inch Monitor
1.) Remove the Monitor's Rear Cover as described below:
a.) Remove the two fixing screws
b.) Remove the Monitor's Rear Cover.
1
EVISION
Table 9-9
on page 9-8.
Figure 8-32 Monitor Rear Cover's fixing screews
Chapter 8 - Replacement Procedures
V
S60N/V
IVID
(Figure
8-32).
S70N
B
S
IVID
ASIC
ERVICE
Power Shut Down on page
M
ANUAL
8-31