1 If the analyzer fails the channel A compression test, suspect the A5 A sample
2 Suspect the A6 B sampler assembly if the analyzer fails the channel B
Required
Equipment
Procedure
1 Connect the equipment as shown in
Performance Tests – Option 011

16. Source and Receiver Harmonics (Option 002 Only)

These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
The power cord is connected to internal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
assembly. Repeat this test. Replace the sampler assembly if the problem still
exists.
compression test. Repeat this test. Exchange the sampler assembly if the
failure still occurs.
16. Source and Receiver Harmonics (Option 002 Only)
Perform this test to determine the 2nd and 3rd harmonics of the source,
receiver, and the source/receiver combination.
2-way Power Splitter, 50Ω . . . . . . . . . . . . . . . . . . . Agilent 11667A Option 001
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Attenuator, 30 dB . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 030
RF cable set, 50Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11851B