Repair and Replacement Procedures

Removing and Replacing the A15 CPU Board

Removing and Replacing the A15 CPU Board
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-12
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. Disconnect the hard disk drive assembly (HDDA) ribbon cable and interface board from
the A15 CPU board by removing the two screws (item
4. Disengage the A6 SPAM board from the A15 CPU board by performing the following
steps.
a. Lift the two extractors located at each end of the board.
b. While holding onto the extractors, lift the board halfway out of the slot.
c. Let the extractors drop to their normal position.
d. Lower the board into the slot.
The board should now rest above its normal seated position.
5. With a T-10 TORX driver, remove the 14 screws (item
A15 CPU board.
6. Gently pull on the finger grip (item
analyzer.
7. Slide the A15 CPU board out of the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-28
for this procedure.
) to disengage the A15 CPU board from the
Table 7-2 on page
PNA Series Microwave Network Analyzers
"Removing the Covers" on page
).
) from the rear panel of the
7-66.
Service Guide E8361-90001
E8361A
7-6.