Required
Equipment
Procedure
1 Connect the equipment as shown in

2. Source Power Range, Linearity, and Accuracy

Perform this test to verify the analyzer's power range and linearity at different
CW frequencies throughout the analyzer operating frequency range.
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
2-way Power Splitter (2), 50Ω . . . . . . . . . . . . . . . .Agilent 11667A Option 001
Attenuator (2), 20 dB . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 020
RF Cable Set, 50Ω, Type-N. . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
Path loss calibration
Figure 6-2. Path Loss Calibration Test Setup (#1)
Performance Tests – Option 011
2. Source Power Range, Linearity, and Accuracy
Figure
6-2.
6-5