Frequency Readout Accuracy
Test Limits
Frequency Readout Accuracy is equivalent to the following equation:
± (0.25% x span + 5% X RBW + 2 Hz + 0.5 x horizontal resolution)
NOTE
See results table for actual values.
Test Description
The frequency readout accuracy is measured in several spans and center frequencies that
allow both internal analyzer synthesizer modes and prefilter bandwidths to be tested.
Frequency reference error is eliminated by using the same frequency standard for the
analyzer and signal source.
Item
Adapter
Type-N(m), to 3.5 mm(f)
Adapter
3.5 mm(f) to 3.5 mm(f)
Cable
APC 3.5 mm, 1 meter
Cable
BNC, 120 cm
Synthesized
Frequency: Capable of 2 GHz (must have external
Sweeper
reference input)
Figure 3-4. Frequency Readout Accuracy Test Setup
Chapter 3
Critical Specification
(for this test)
PSA Functional Tests
Frequency Readout Accuracy
Recommended
Agilent Model
1250-1744
5061-5311
8120-4921
10503A
83630B