Frequency Accuracy Test; Description; Test Equipment - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

2. FREQUENCY ACCURACY TEST

Description

This test checks the frequency accuracy of the E5071C test port output signal. The
frequency accuracy is checked at 50 MHz and 8.5 GHz with a frequency counter. Since the
E5071C employs a PLL frequency synthesizer for the signal source, the frequency
accuracy test at these two frequency points can verify the accuracy for the entire frequency
range.

Test equipment

Frequency Counter
48 inch BNC cable
N(m)-BNC(f) adapter
Chapter 2
2. FREQUENCY ACCURACY TEST
Agilent 53181A with Opt. 010 and 124 or Agilent
53131/2A with Opt. 010 and 124
Agilent p/n 8120-1840
Agilent p/n 1250-0780
Performance Test
31