Repairing And Recovering Option Data; Recovery Of Data After Repair; Recovery Of Data If Option Or Model Numbers Are Incorrect - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Option Enable Utility

Repairing and Recovering Option Data

Repairing and Recovering Option Data
License, option, and model number data are stored in an EPROM on the A10 frequency
reference board and written into the gen.lic file on the hard disk drive. If the data stored
in either of these items is lost, it can be recovered from the other item.
If the data on both the hard disk drive and the frequency reference board is lost, it can not
be recovered. Contact Agilent for assistance. Refer to

Recovery of Data After Repair

• If the A10 frequency reference board is replaced, use the
that has been lost as a result of the repair:
1. Select
from the
Repair
2. Click the
Freq Ref board has been replaced
3. Click
Begin Repair
• If the hard disk drive is replaced, a new gen.lic file is automatically created when the
application starts, by retreiving the data from the EPROM.
Network Analyzer

Recovery of Data if Option or Model Numbers are Incorrect

If the analyzer option or model numbers are not listed correctly on the analyzer display, in
either the Option Enable Dialog Box or the About Network Analyzer display, you should
regenerate the options license file, gen.lic:
1. Exit the
Network Analyzer
2. Remove (or rename) the existing gen.lic file:
a. Open
Windows Explorer
Analyzer.
b. Delete (or rename) gen.lic.
3. Exit the
Windows Explorer
4. Open the
Network Analyzer
file when it starts.
5. Check the option listing:
a. On the
System
b. On the
menu, click
Help
6. If the options are still not listed correctly, contact Agilent for assistance. Refer to
"Contacting Agilent" on page
B-4
Option Enable
. The data is written from the gen.lic file into the EPROM.
application.
and navigate to C:\Program Files\Agilent\Network
application.
application. The application will generate a new gen.lic
menu, point to
Service
About Network Analyzer
2-9.
PNA Series Microwave Network Analyzers
"Contacting Agilent" on page
Repair
menu bar (see
Figure
check box.
, and then click
Option Enable
.
E8361A
2-9.
selection to recover data
B-1).
, or
Service Guide E8361-90001