Removing the IonSABRE II probe

Required materials
Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe:
Warning:
source can be contaminated with biohazardous and/or toxic
materials. Always wear chemical-resistant, powder-free gloves
while performing this procedure.
Warning:
prepared for working on the source before commencing this procedure.
1.
Prepare the instrument for working on the source (see
2.
Remove the corona pin (see
3.
Disconnect the diverter valve tubing from the IonSABRE II probe.
4.
Unscrew the probe locking ring.
5.
Carefully remove the probe from the probe adjuster assembly.
3-10
Changing the Mode of Operation
The LC system connections, IonSABRE II probe, and
To avoid electric shock, ensure that the instrument is
page
4-15).
page
4-8).