Removing The Probe Adjuster Assembly Probe And Source Enclosure Seals - Waters Xevo G2-XS QTof Overview And Maintenance Manual

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Renew the following seals:
• Probe adjuster assembly probe seal
• Probe adjuster assembly nebulization gas seal
• Source enclosure seal
• Nebulizer gas seal
• Desolvation gas seal
7.31.1 Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
• Chemical-resistant, powder-free gloves
• Protective eyewear
• O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
1. Remove the source enclosure from the instrument (see
from the instrument (Page
2. Use the O-ring removal kit (see
remove the probe and nebulizer gas seals from the probe adjuster assembly.
91)).
Removing O-rings and seals (Page
March 31, 2021, 715004496 Ver. 04 (previously released as Rev. D)
Removing the source enclosure
Page 210
99)) to carefully