Clean or Replace the Collecting Electrode (FID on ECD)
OPERATING SEQUENCE
Clean or Replace the Collecting Electrode (FID on ECD)
Materials needed
for replacement:
for cleaning:
WARNING! Any maintenance or repair operations involving the radioactive source of the ECD
must be performed ONLY by qualified personnel duly authorized to handle
radioactive material. Please contact your local Thermo Fisher Scientific Technical
Service office.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detectors must be cooled to room temperature.
PRECAUTIONS
Maintenance and Troubleshooting Manual
Screwdriver
2.5 mm Allen wrench
3 mm Allen wrench
Tweezers
Collecting electrode (P/N 277 003 50)
Ultrasonic cleaning bath
GC-grade hexane
GC-grade toluene
Fine emery paper (optional)
Maintaining Detectors in Stack Configuration
Reagents Safety Information
Chapter 17
for more information.
349