Clean or Replace the Liner
18. Clean the septum support from possible
fragments left by the septum and reinsert it into
the injector
19. Place the septum holder on the top of the
septum support.
20. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
21. Press LEFT CARRIER or RIGHT CARRIER,
depending on which injector is operating, scroll
to
and press ON.
Pressure
22. Press OVEN, scroll to
23. Press LEFT INLET or RIGHT INLET, depending on
which injector is operating, scroll to
press ON.
24. Perform a leak check according to
Leak Check (Capillary Column)
Sequence on page 181.
Maintenance and Troubleshooting Manual
and press ON.
Temp
Temp
Performing a
Operating
and
Maintaining a Split/Splitless Injector
Chapter 2
39