ECD Troubleshooting

Troubleshooting the ECD

Symptom
High base frequency
Maintenance and Troubleshooting Manual
Table 12-3. ECD Troubleshooting Table
Diagnosis
Impure gas supply.
No (or insufficient) flow of
make-up gas.
Excess column bleeding.
Leaks on carrier and/or make-up
gas lines.
Contaminated collecting electrode
Chemically contaminated
radioactive source.
Pulse width not correctly set.
Maintaining an Electron Capture Detector
Remedy
Use high purity grade gases and
filters to trap moisture and oxygen.
Increase make-up gas flow rate.
Column conditioning is needed.
Run a leak test.
Refer to the
Performing a Leak
Check (Capillary Column)
Operating Sequence on page 181.
for instructions.
Clean the collecting electrode, as
described in the
Clean the
Collecting Electrode
Sequence on page 226, or replace it.
Refer to the notes reported in the
ECD Maintenance
paragraph on
page 222.
Contact your customer support
organization. Appendix B,
Customer
Communication, contains
contact information for Thermo
Fisher Scientific offices and
affiliates worldwide.
to 1.0 s for
Set
Pulse width
Nitrogen and 0.1 s for Ar/CH