Replace the Collecting Electrode
11. Using forceps, reconnect the gold contact pin.
12. Remount the connector holder assembly
verifying the contact efficiency.
13. Place the detector cell on the detector base body
and tighten the fixing screw.
14. Screw the detector cap into its housing.
15. Reconnect the signal and ignition cables to the
detector.
16. Reset the detector to the required operating
conditions. Refer to the relevant chapter of the
TM
TRACE
GC Ultra Operating manual for
more information.
Maintenance and Troubleshooting Manual
Maintaining a Nitrogen Phosphorus Detector
Chapter 13
259