Clean the Collecting Electrode

OPERATING SEQUENCE
Clean the Collecting Electrode
Materials needed
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
Maintenance and Troubleshooting Manual
Screwdriver
Pliers
Forceps or tweezers
Ultrasonic cleaning bath
Liquid detergent
Clean compressed air
GC-grade methanol
Clean paper towel
Maintaining a Nitrogen Phosphorus Detector
Reagents Safety Information
Chapter 13
for more information.