Chapter 9
Maintaining a PTV Injector
24. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector cap. You could damage the septum and affect
performance.
25. Press RIGHT CARRIER, scroll to
press ON.
26. Press OVEN, scroll to
27. Press RIGHT INLET, scroll to
28. Perform an automatic leak check according to
Performing a Leak Check (Capillary Column)
Operating Sequence on page 181.
164
Pressure
and press ON.
Temp
and press ON.
Temp
and
Maintenance and Troubleshooting Manual
Replace a Broken Liner