Adjust a Quartz Wool Packing
9. Use the screwdriver provided with the GC to
remove the injector liner cap.
10. Use tweezers to remove the liner with the
graphite seal.
11. Remove the graphite seal from the liner. Use
tweezers to handle the liner.
CAUTION
Be careful not to damage the
graphite seal or allow graphite
to entering the liner. Should this
occur, clean the liner with an
inert gas.
PRECAUTIONS