Platform Handle
Adjustment

Control Panel

Clean/Replace
Condenser Air Filter
3-2
MaxQ SHKE8000 Series Shakers
1. Grasp the platform handle and pull up on the inner release bar to
move the handle to a horizontal (down) position.
2. Remove the screws from the sides of
the handle, as indicated in Figure 3-
2. Remove the handle.
3. Turn the adjustment screws just
one-half turn clockwise to increase
the clamping force, or one-half turn
counterclockwise to reduce the
clamping force.
4. Reassemble the handle and install the side screws. Slide the platform
into the shaker chamber. Place the handle into a vertical position.
5. Close the door, and with the unit empty, set the speed to 300 RPM. If
there is undue vibration noise, repeat the steps above. If not, return the
unit to service.
The control panel uses sealed push buttons and liquid crystal display. It
may be cleaned with a mild detergent and dried with a soft cloth.
The condenser air filter is located behind the grille on the front of the
refrigeration assembly attached to the side of the shaker (Model
SHKE8000-7 & SHKE8000-8CE). The grille is secured by four press-in
type retainers at the top and bottom. Remove it by grasping the edge of
the cover and pulling outward.
The air filter is held in place with a flange
around its perimeter (Figure 3-3) and is
easily removed. It may be replaced, or
washed in water with a mild detergent and
dried between two lint-free towels.