Replace the Collecting Electrode
OPERATING SEQUENCE
Replace the Collecting Electrode
Materials needed
•
•
•
•
WARNING! Any maintenance or repair operations involving the radioactive source must be
performed ONLY by qualified personnel duly authorized to handle radioactive
material. Please contact your local Thermo Fisher Scientific Technical Service office.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
Maintenance and Troubleshooting Manual
Screwdriver
2.5 mm Allen wrench
Tweezers
Collecting Electrode (P/N 277 003 50)
Maintaining an Electron Capture Detector
Reagents Safety Information
Chapter 12
for more information.
229