Chapter 3
Maintaining a Large Volume Splitless Injector
12. Using tweezers, remove the liner from the bath
and dry it with compressed clean air.
NOTE
For trace analysis, you should pre-
treat the liner with a suitable
silylating reagent prior to re-
inserting it into the injector.
PRECAUTIONS
13. Using tweezers, insert the liner into the injector
and push it gently towards the bottom fitting.
CAUTION
Make sure you insert the
bevelled end of the liner
towards the bottom of the
injector.
64
Clean or Replace the Liner
Maintenance and Troubleshooting Manual